Page C-4 P/N. 12457-001 Rev. 2
Power System User Guide
Note: These requirements provide reasonable protection against harmful electromagnetic interference in
a typical medical installation. However, high level of radio-frequency emissions from electrical devices,
such as cellular phones, may disrupt the performance of this device. To mitigate disruptive
electromagnetic interference, position this device away from radio frequency transmitters and other
sources of electromagnetic energy.
C.4 FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
RF electromagnetic fields
immunity
Meets IEC 61000-4-3, Level 2
EFT/burst Meets IEC 61000-4-4, Level 2
Surge immunity Meets IEC 61000-4-5, Level 2
Conducted RF disturbances
immunity
Meets IEC 61000-4-6, Class A
Dips, interruptions, and
variations
Meets IEC 61000-4-11
Harmonics current emissions Meets IEC 61000-3-2, Class A
Radiated emissions Meets CISPR 11/EN55011, Group 1, Class A
FCC part 15, Class A
Safety Meets IEC/EN60601-1
Table C-3 Multi-Chemistry Battery Charger Specifications (Page 2 of 2)