Page 5-6 P/N. 12457-001 Rev. 2
Power System User Guide
5.5 Battery Rotation
After every use, at the beginning of a shift, or at least once every 24 hours, the Battery in the AutoPulse
should be replaced with a fully charged Battery.
Figure 5-5 Four Battery rotation
Four Battery Rotation—to be done after every use and/or once per shift (see Figure 5-5 above).
Remove the Battery from the AutoPulse, and place it into the Charger.
Remove the spare Battery, and place it into the Charger.
Take two fully charged Batteries from the Charger, check for green LEDs on each Battery, and then
place one into the AutoPulse, and use the second as a spare.
Power on the AutoPulse and ensure no faults are displayed.
1
Replace the spare Battery
with a freshly-charged
Battery from the
Charger.
3
Charge the Platform Battery
and the Spare Battery.
Check the Battery
status
LED
s before
placing a Battery
in service.
2
Replace the Platform
Battery with a freshly
charged Battery from
the Charger.
Recommended daily 4 Battery maintenance routine